The Effect of Kilovoltage and Milliampere-Second Parameters on CT Number: Study Phantom Quality Control CT Scan

Muhammad Irsal - Nama Orang; Guntur Winarno - Nama Orang; Gando Sari - Nama Orang; Aulia Narindra Mukhtar - Nama Orang; Aulia Narindra Mukhtar3 - Nama Orang;

CT number is one of the indicators in determining the diagnosis of disease, so it is necessary to ensure the level of accuracy. The aim is to determine the parameters of kilovoltage and tube current to the CT value. Descriptive quantitative analytic research method with previously tested linearity and the suitability of the CT number. Furthermore, the variation of kilovoltage: 80 kVp, 100 kVp, 140 kVp and the tube current value is 192 mAs. After that, the tube current variation was continued: 80 mAs, 100 mAs, 140mAs and the voltage value was 120 kVp. Phantom image results were analysed using radiant viewer software with a region of interest (ROI) size of 5 mm. Then analyse the correlation coefficient to determine the degree of relationship between the kilovoltage and milliampere-second parameters to the CT number. The results of the linearity test and the suitability of the CT number value were within the tolerance limit. For variations in kilovoltage to CT number, the correlation coefficient values are water R2=0.09, fat R2=1, soft tissue R2=0.9, bone R2=0.7. As for the milliampere-second variation of the CT number, the correlation coefficient values are: water R2=0.7, fat R2=0.8, soft tissue R2=1, bone R2=0.7.


Ketersediaan
#
Kampus A 001.43
ART0000000000545
Tersedia namun tidak untuk dipinjamkan - For Reading
Informasi Detail
Judul Seri
-
No. Panggil
001.43
Penerbit
Jakarta : Poltekkes Kemenkes Jakarta II., 2022
Deskripsi Fisik
-
Bahasa
Indonesia
ISBN/ISSN
-
Klasifikasi
001.43
Tipe Isi
-
Tipe Media
-
Tipe Pembawa
-
Edisi
-
Subjek
milliampere-second
kilovoltage
CT number
Info Detail Spesifik
Publikasi pada SANITAS: JURNAL TEKNOLOGI DAN SENI KESEHATAN ; ISSN : 1978-8843 (PRINT) / 2615-8647 (ONLINE) Vol. 13 No. 2: 237 - 244; https://doi.org/10.36525/sanitas.2022.20
Pernyataan Tanggungjawab
-
Versi lain/terkait

Tidak tersedia versi lain

Lampiran Berkas
  • The Effect of Kilovoltage and Milliampere-Second Parameters on CT Number: Study Phantom Quality Control CT Scan
    Other Resource Link
Komentar

Anda harus masuk sebelum memberikan komentar


NPP 3174072C0000001

Link

  • Web Utama
  • Web Perpustakaan
  • Repository
  • KINK
  • Eresources PNRI

Kontak Kami

Perpustakaan Poltekkes Kemenkes Jakarta II
Jalan Hang Jebat III/F3 Kebayoran Baru Jakarta 12120

Telp.: +6221-7397641
Email: perpus@poltekkesjkt2.ac.id

Cari


Download

Tumbuh Bersama

Tumbuh bersaama kami dengan pengetahuan yang menakjubkan

Temukan, Pelajari, Berkembang

Perpustakaan, Membangkitkan Keajaiban Lintas Generasi

Nyalakan Masa Depan Anda

Alat dan Panduan untuk Perjalanan Setiap Pembelajar

Baca. Jelajahi. Raih

Dimana Imajinasi Menjadi Kenyataan


Pilih subjek yang menarik bagi Anda
  • Karya Umum
  • Filsafat
  • Agama
  • Ilmu-ilmu Sosial
  • Bahasa
  • Ilmu-ilmu Murni
  • Ilmu-ilmu Terapan
  • Kesenian, Hiburan, dan Olahraga
  • Kesusastraan
  • Geografi dan Sejarah
Icons made by Freepik from www.flaticon.com
Pencarian Spesifik