Analysis Of Rooftop Skyshine Radiation Exposure With Angle Of Gantry Linear Accelerator 180° In Radiotherapy Unit Of Pertamina Central Hospital Jakarta
ABSTRACT
Principally, the measurement of skyshine radiation exposure is important to be conducted if there is public
activity in a certain place in which the source of radiation can emit its scattering radiation to all directions
including to the rooftop of a Radiotherapy building. the Regulation of the Head of the National Nuclear
Power Energy No. 8 Year 2011 Article 32a and No. 4 year 2013 article 23a namely 1 mSv/year. As for the
design of this research, it is quantitative descriptive by conducting the measurement of radiation exposure to
44 points of measurement, research instrumentations used herein are 2 surveymeter, water phantom, Linac
Primus 2D Plus. Result radiation exposure measurement are 6 point areas with the values above 1 mR/hr.
Each of the average value of the measurement of the exposure rate at each point is calculated by the annual
dose formula. All the final calculation results with the annual dose formula in mSv/year unit yields the values
below 1 mSv/year at the 44 points measurement. the rooftop area of Radiotherapy building through Child
Polyclinic Pertamina Central Hospital is under a safe limit with all calculation results of annual dose values
below 1 mSv/year.
Keywords : Skyshine radiation, Exposure rate, Annual dose