ANALYSIS OF EYE LENS EFFECTIVE DOSE OF WORKERS DURING CARDIAC CATHERIZATION EXAMINATION IN A CATHLAB ROOM

Sriyatun - Nama Orang; Samsun - Nama Orang; Eka Putra Syarif Hidayat - Nama Orang; Guntur Winarno - Nama Orang; Gando Sari - Nama Orang; Ary Sasongko - Nama Orang;

C-Arm equipment is commonly used for three-dimensional patient imaging in real time using fluoroscopy techniques. This study aims to evaluate the effective dose of eye lens received by radiation workers during cardiac catheterization. The main tool uses the C-Arm equipment and the calibrated TLD detector chip. The object of observation is the eye lens of radiation workers consisting of one radiologist and one nurse. Catheterization was carried out to 10 patients with different examination times. As the observation results, for the eye lens without protective glasses; the average effective doses received by medical doctors and the nurse for a single imaging are, respectively in the ranges of 0.0011 and 0.0054 mSv, and in the ranges of 0.0010 mSv and 0.0025 mSv, or about 0.0037 mSv/h - 0.0135 mSv/h for the medical doctors and about 0.0027- 0.00830 mSv/h for the nurse. Estimated dose received by medical doctors and nurse are, respectively, in the range of 0.0037 - 0.0135 mSv/h and 0.0027-0.0083 mSv/h


Ketersediaan
#
Kampus A 001.43
ART0000000000576
Tersedia namun tidak untuk dipinjamkan - For Reading
Informasi Detail
Judul Seri
-
No. Panggil
001.43
Penerbit
: JSTNI., 2022
Deskripsi Fisik
-
Bahasa
Indonesia
ISBN/ISSN
-
Klasifikasi
001.43
Tipe Isi
-
Tipe Media
-
Tipe Pembawa
-
Edisi
-
Subjek
Fluoroscopy
C-Arm
cardiac catheterization
effective dose
eye lens dose
Info Detail Spesifik
Publikasi pada Jurnal Sains dan Teknologi Nuklir Indonesia Vol 23, No. 1 (2022) 32 – 35; DOI : 10.17146/jstni.2022.23.1.6753
Pernyataan Tanggungjawab
-
Versi lain/terkait

Tidak tersedia versi lain

Lampiran Berkas
  • ANALYSIS OF EYE LENS EFFECTIVE DOSE OF WORKERS DURING CARDIAC CATHERIZATION EXAMINATION IN A CATHLAB ROOM
    Other Resource Link
  • Turnitin : Analysis of Effective Radiation Dose in Cathlab Room At The Time of Insertion of Cardiac Catheterization
Komentar

Anda harus masuk sebelum memberikan komentar


NPP 3174072C0000001

Link

  • Web Utama
  • Web Perpustakaan
  • Repository
  • KINK
  • Eresources PNRI

Kontak Kami

Perpustakaan Poltekkes Kemenkes Jakarta II
Jalan Hang Jebat III/F3 Kebayoran Baru Jakarta 12120

Telp.: +6221-7397641
Email: perpus@poltekkesjkt2.ac.id

Cari


Download

Tumbuh Bersama

Tumbuh bersaama kami dengan pengetahuan yang menakjubkan

Temukan, Pelajari, Berkembang

Perpustakaan, Membangkitkan Keajaiban Lintas Generasi

Nyalakan Masa Depan Anda

Alat dan Panduan untuk Perjalanan Setiap Pembelajar

Baca. Jelajahi. Raih

Dimana Imajinasi Menjadi Kenyataan


Pilih subjek yang menarik bagi Anda
  • Karya Umum
  • Filsafat
  • Agama
  • Ilmu-ilmu Sosial
  • Bahasa
  • Ilmu-ilmu Murni
  • Ilmu-ilmu Terapan
  • Kesenian, Hiburan, dan Olahraga
  • Kesusastraan
  • Geografi dan Sejarah
Icons made by Freepik from www.flaticon.com
Pencarian Spesifik